Inspection #: 874145
Inspection Type: Final Electrical
Inspector: M.Green
Result: Correction
Comment: Chiller to serve linear accelerator installed on rooftop.

Inspection Corrections
DateStatusCodeDescriptionRefLocationComments
10/2/2025TransferCorrect Below* Drawings.Show 15A fused disconnect. Contractor has installed a 15HP rated snap switch. Engineer to change drawings as discussed.
10/2/2025TransferCorrect Below* Chiller ratings.Approved drawings show chiller with MCA rating of 14.5A and using an existing 20A, 3-pole breaker. Chiller nameplate shows an MCA rating of 17.6A and MOP of 25A. Engineer needs to determine adequacy of circuit and change drawings as discussed.
10/2/2025TransferCorrect Below* Chiller unit listing.Unit does not appear to have a listing label. Unit will need to be replaced or field evaluated if a valid listing label is not located on the unit.