Date | Status | Code | Description | Ref | Location | Comments |
10/20/2022 | Corrected | Correct Below | * | | Therapy room | Receptacles on north wall have no power. |
10/20/2022 | Corrected | Correct Below | * | | Ceiling | Jboxes need cover plates. |
10/20/2022 | Corrected | Correct Below | * | | X-ray room | Receptacle next to 2 gang blank has no power. |
10/12/2022 | Corrected | Correct Below | * | | Panel | Update panel schedule, make up is incomplete. |
10/12/2022 | Corrected | Correct Below | * | | Throughout | MC cable needs to be supported so it is not lying on suspended ceiling. |
10/12/2022 | Corrected | Correct Below | * | | Patient care areas | All branch circuits serving these spaces shall be provided with an effective ground fault current path. 2017 NEC 517.13 A
This includes circuits that go to light switches as well as receptacles. |