Inspection #: 386240
Inspection Type: Final Electrical
Inspector: H8
Result: Correction
Comment: Request an AM final inspection, please contact Walter Ruark on site, 538-5579, thanks, Kevin Hunter Requested by: Kevin Hunter 272-3006 khunter@encoreak.com ||

Inspection Corrections
DateStatusCodeDescriptionRefLocationComments
1/27/2016TransferEF1Correct Below in break room kitchenall 120V 15-20 amp recetpacle must be GFI protected and readily accessible
1/27/2016TransferEF1Correct Below at water heater disconnectbond around knockouts in disconnect
1/27/2016TransferEF1Correct Below several exam roomsreceptacles within 6' of sinks must be GFI protected
1/27/2016TransferEF1Correct Below in sink in room off of nurses station room 218all receptacles within 6' of sink must be GFI protected